Detector sensitive area
1800 cm2 ( 2 x windows 300 x 300 mm ea.)
30cm (12”) depth x 30cm (12”) width
Chamber entrance height = 2.2cm (.875") tall
Detector Type
Gas Flow
Proportional
Gas Used
P-10 Nuclear Counting Gas
90% Argon, 10% Methane
(DOT Class 2.2 Nonflammable Gas)
Background
Less than or equal to 0.006 alphas/cm2/hr
Detection Efficiency
Greater than 80% (2 pi )
Lower Limit of Detection (LLD)
Bkgd of 6 = 0.001 @ 40 hrs, 0.0006 @ 100 hrs
Bkgd of 3 = 0.001 @ 20 hrs, 0.0004 @ 100 hrs
Operating Voltage
+ 800 VDC
Alpha Detector Window
250 µg/cm2 metalized Mylar
Overall Outside Dimensions
46cm (18”) wide x 55cm (21.5”) deep
16.5cm (6.5”) tall
Weight
9 kg (20 lbs.)
Electronics
Detector Bias Power Supply, 1Kv
On board Charge Sensitive Pre-amp, Linear Amplifier, Integral Discriminator.
Gas Handling
Dual Stage Gas Regulator
Adjustable Flow Controller
Pricing and Delivery
Please contact us
for current pricing and delivery schedules.
*Specifications subject to change without notice.
copyright Alpha Sciences Inc. 1999-2009